|
In ASTM C511-04, it talks about
calibration of the recording thermometer, then continues to
discuss how to perform verification of the recording thermometer.
If the recording thermometer is calibrated every six months,
what logic says we have to perform verification? This seems
redundant. Can you give us a little insight of this logic?
Excerpt from Section 5.1, ASTM
C511, Standard Specification for Mixing Rooms, Moist Cabinets,
Moist Rooms, and Water Storage Tanks Used in the Testing of
Hydraulic Cements and Concrete:
"The recording thermometer
shall be calibrated at least every six months or whenever there
is a question of accuracy.
Perform the verification of
the recording thermometer by comparing the temperature reading
of the recording thermometer with the temperature reading of
a reference thermometer during the normal operation of the moist
cabinet or moist room.
The thermometer used as the
reference thermometer must be accurate and readable to 0.5°C."
-- Unknown
|